Advanced internal calibration of the leak
detector
3
/3
GB 00970 -
Edition
05 -
June
08
Checking
function with an
external calibrated leak
It is possible to perform a calibration checking with an external
calibrated leak (position parameter = external).
But
it is not automatic
and in the case, the leak
detector performs
a
complete
calibration (not
only a comparison like with the internal calibrated leak).
When the first of the 2
counters
is reached, a
“i“
will flash at the right
end of the
1st
line of the LCD.
READY
FOR CYCLE i
...
...
...
The
“i”
flashing will stay on the
LCD
until operator
starts
an external
calibration.
C
303
An audio
message advises
the operator in the
process
to follow
.
Note: It is still possible to start a
test
cycle even without performing an
internal calibration
.
On
request
calibration
checking
C
301