Adixen ASM GRAPH D+ - C 550 - Memo function

Default Icon
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C
550
Memo function
1/
1
Test
mode
mba
r
.l/s
Gross leak
1x10
-9
to
1
Normal
1
x
10
-
11
to
3.10
-4
Sniffing
1x10
-7
to
1.10
-
1
GB 01266 -
Edition
06 -
June
08
Zero activated
Bargraph zoom on reject point
not activated
+
or
Bargraph zoom on reject point
activated
LEAK RA
TE
COR
=>
LEAK RA
TE
COR
LEAK RA
TE
COR
=>
LEAK RA
TE
COR
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
10
-3
10
-2
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
10
-3
10
-2
Reject
point
0.0E-00 displayed
only for 1 s, then
1
/
100
leak value
displayed
(*)
Example: displayed
value
=
1
x
10
-
10
mbar
.l/s
10
-2
10
-3
10
-4
10
-5
10
-6
0.0E-00 displayed
only for 1 s, then
1
/
100
reject point
value displayed
(*)
Example: reject point
=
1x10
-7
mbar
.l/s
Displayed leak
=
10
-2
10
-3
10
-4
10
-5
10
-6
-7
Reject
point
10
-7
10
-8
1x10
-9
mbar
.l/s
10
10
-8
10
-9
10
-9
10
-1
0
10
-10
10
-1
1
10
-11
10
-1
2
mbar
.l/s
10
-12
mbar
.l/s
(*) Note:
The
1
/
100
reject point value or
1
/
100
leak value
displayed is limited to the low limit of t
he
measurment
range in the
selected test
mode.
Example: ASM 142 in Normal
test
mode
The
1
/
100
reject point value or
1
/
100
leak value will never be
lower than
1
x
10
-
11
mbar.l/s.
Refer
to A 800 for the limit
values
according to the leak
detector.