Needle replacement
(continued)
With
needlenose
pliers, take out the original needle.
Put
the new needle (ref. 1 ) and cut to the
same
length as the
original
needle.
Note:
For the flexible sniffer probe
(SNCxE3Tx
and
SNCxE4Tx),
the needle
length should be 2.5
cm/0.98
inch (ref. 2 ).
Bend the needle as shown in the picture
3
.
Put
the needle in the nipple, the crooked end to outside (ref. 4 ) and
push
it in with a 2.5 allen key to stop.
Put
back the filter.
Needle
Needle for
flexible sniffer probe
1 2 3
4
Key
O'ring
installation
When you
unscrew
the nipple from the probe,
it is possible for the o’ring to come out with the
nipple : you should put it back (ref.
1 ).
Place
the o’ring on the nipple of the aspiration
tube in the sniffer body (ref.
2 ).
Push
the o’ring with the nipple of the
sniffer
probe (ref. 3 ).
1
Nipple
O'ring
Body
The
o’ring is correctly placed on the nipple (ref. 4 ).
Screw on the nipple of the sniffer probe.
2
3
4
7
/8