A help for control panel utilization/access.
Operating principle of the control panel
C
110
Access
to level 4 - Password
C
130
Setting
and
maintenance
part
presentation
of the control panel
Access
to
parameters
and
parameters
active
depending on authorization
C
120
Summary
of screens
Complete displays list with
access
way and
associated
sheet
C
140
Purpose
The
helium pollution prevention function
prevents
the
detector
from
being polluted by helium when the part or installation to be tested
contains
a lot of helium.
Leak detector
in
test
mode, if the signal
increases
quickly above
depollution reject point, the cycle is automatically ended and the leak
detector stays
in
stand-by
mode until the helium has decreased.
Note:
There
is no
message
to inform the operator that the
detector
is
over again ready for a new test.
Note: