Adixen ASM GRAPH D+ - Activate;Deactivate the Helium Pollution Prevention; Date - Time - Language - Unit

Default Icon
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C
570
Date -
Time
-
Language
- Unit
1/
1
GB 00270 -
Edition
04 -
June
08
A help for control panel utilization/access.
Operating principle of the control panel
C
110
Access
to level 4 - Password
C
130
Setting
and
maintenance
part
presentation
of the control panel
Access
to
parameters
and
parameters
active
depending on authorization
C
120
Summary
of screens
Complete displays list with
access
way and
associated
sheet
C
140
Purpose
The
helium pollution prevention function
prevents
the
detector
from
being polluted by helium when the part or installation to be tested
contains
a lot of helium.
Leak detector
in
test
mode, if the signal
increases
quickly above
depollution reject point, the cycle is automatically ended and the leak
detector stays
in
stand-by
mode until the helium has decreased.
Note:
There
is no
message
to inform the operator that the
detector
is
over again ready for a new test.
Note:
Helium pollution prevention
function not activated
+
Big quantity of Helium in
the pumped parts
=
Detector
polluted
Depollution
= Wait and pump a few hours (the best
solution is one night)
=
Purge
with N2 flow or air
flow
Note: When the
detector reaches
the 10
-4
mbar
threshold,
it
automatically pass in
gross
leak
test
mode.
Activate/Deactivate
the Helium pollution
prevention
2x
Adjustment
Adjustment
C
110
off
on