C
580
Fault
/
information indicator and display
3
/6
SERVICE
INFORMATION
new
filament
required
PRESS
RESET
GB 02465
-
Edition 03
-
September
06
Information
2 information types: user and
service information
.
W
arning:
-
no indicator on the digital display
-
on the
LCD,
a
“i”
flashing at the right end of the 1
st
line.
LEAK
RA
TE
COR
...... i
......
......
......
- The
digital voice
advises
the operator of the procedure to follow
.
Result:
Doesn’t
affect the
functions
of the leak detector
-
User
information
Only an indication that the leak
detector
is in a particular
status
which
may require an action from the
user
in order to return to a standard
situation
-
Service
information
Only an indication that the leak
detector requires
a
service
or
maintenance
action.
Message:
To
read the
messsage
NEXT
A clear
message describes
the fault on line 2 and
3.
User
information display
Service
information display
USER
INFORMATION
filament request
off
PRESS
RESET
Duration:
After display of the clear
message,
the indicator and the clear
message
are erased but they will be reactivated at the next
start-up
of
the leak
detector
or at each unauthorized
request
by the
user
or
30 min later, if the origin of the
message
is still present.
Remedy:
-
User
information
Can be eliminated by an action which is
accessible
by the user.
-
Service
information
Can be eliminated by a
service
action on the involved
component
and
by
resetting
the corresponding configuration parameter.
This
is only
accessible
by the
customer
service.
Service
instr
uctions
E