GPIB OPERATION
3-2-32
SWEEP? (continued)
:CFRQ?
Prepares message containing information on Carrier Frequency
Sweep settings in the following format:
:SWEEP:CFRQ:START <nr2>;STOP <nr2>;STEP <nr1>;TIME <nr1>
Sample response:
:SWEEP:CFRQ:START 1230000.0;STOP 1330000.0;STEP 100;TIME
20
:VALUE?
Prepares message containing information on the current carrier
frequency marker settings in the following format:
:SWEEP:CFRQ:MKRNUM <nr1>;VALUE <nr2>;<status>
where: <status> is a program mnemonic indicating whethe
the selected Marker is ON or OFF.
Sample response:
:SWEEP:CFRQ
:MKRNUM 2;VALUE 1240000.0;MKRON
:RFLV?
Prepares message containing information on RF Level Sweep
settings in the following format:
:SWEEP:RFLV:START <nr2>;STOP <nr2>;STEP <nr1>;TIME <nr1>
Sample response:
:SWEEP:RFLV:START -107.0;STOP -27.0;STEP 80;TIME 50;
:VALUE?
Prepares message containing information on the current RF
level marker settings in the following format:
:SWEEP:RFLV:MKRNUM <nr1>;VALUE <nr2>;<status>
where: <status> is a program mnemonic indicating whethe
the selected Marker is ON or OFF.
Sample response:
:SWEEP:RFLV:
MKRNUM 2;VALUE -97.0;MKRON
:LFGF?
Prepares message containing information on LF Generator
Frequency Sweep settings in the following format:
:SWEEP:LFGF:START <nr2>;STOP <nr2>;STEP <nr1>;TIME <nr1>
Sample response:
:SWEEP:LFGF:START 300.0;STOP 3000.0;STEP 2700;TIME 1
:VALUE?
Prepares message containing information on the current LF
generator frequency marker settings in the following format:
:SWEEP:LFGF:MKRNUM <nr1>;VALUE <nr2>;<status>
where: <status> is a program mnemonic indicating whethe
the selected Marker is ON or OFF.
Sample response:
:SWEEP:LFGF:MKRNUM 2;VALUE 400.0;MKRON
:LFGL?
Prepares message containing information on LF Generator
Level Sweep settings in the following format:
:SWEEP:LFGL:START <nr2>;STOP <nr2>;STEP <nr1>;TIME <nr1>
Sample response:
:SWEEP:LFGL:START 1.0;STOP 120.0;STEP 120;TIME 10