REMOTE OPERATION AM COMMANDS
4-53
[SOURce][:MODulation]:AM[1]|2:EXTernal:SENSitivity
Description: Selects the sensitivity of the external source input for AM — 1 V RMS or 1 V peak.
Parameters: <CPD>
Valid values: VRMS | VPK
*RST sets: VRMS
[SOURce][:MODulation]:AM[1]|2:EXTernal:SENSitivity?
Description: Queries the sensitivity of the external source input for AM.
Parameters: None
Response: <CRD>
Returned values: VRMS | VPK
[SOURce][:MODulation]:AM[1]|2:INTernal:FREQuency[:FIXed]
Description: Sets the internal AM frequency.
Parameters: <numeric_value>
Valid values: <NRf>(Hz) | MAXimum | MINimum | UP | DOWN | RETurn | REFerence
Set by value, to maximum or minimum, stepping up or down, returning to the last full
setting, or setting the current value to the last full setting.
*RST sets: AM1 = 1 kHz, AM2 = 400 Hz
[SOURce][:MODulation]:AM[1]|2:INTernal:FREQuency[:FIXed]?
Description: Queries the internal AM frequency.
Parameters: None
Response: <NR2>
Returned values: AM frequency in Hz