REMOTE OPERATION SYSTEM COMMANDS
4-165
SYSTem:COMMunicate:REMote
Description: Selects the remote operation interface.
This command is only actioned once the EOM at the end of the message has been
received and all outstanding query responses have been read.
Parameters: <CPD>
Valid values: GPIB | RS232 | ETHernet
*RST sets: No effect
SYSTem:COMMunicate:REMote?
Description: Returns the remote operation interface that the instrument uses.
Parameters: None
Response: <CRD>
Returned values: GPIB | RS232 | ETH
SYSTem:COMMunicate:SERial:BAUD
Description: Sets the baud rate of the serial interface.
This command is only actioned once the EOM at the end of the message has been
received and all outstanding query responses have been read.
Parameters: <numeric_value>
Valid values: 300 | 600 | 1200 | 2400 | 4800 | 9600
*RST sets: No effect on the set baud rate.
SYSTem:COMMunicate:SERial:BAUD?
Description: Returns the baud rate of the serial interface.
Parameters: None
Response: <NR1>
Returned values: 300 | 600 | 1200 | 2400 | 4800 | 9600