GENERAL INFORMATION
1-20
Relative level: 60 dB
DIFFERENTIAL IQ OUTPUT
(OPTION 009)
When differential IQ outputs are enabled, the signal generator RF is CW only.
Output impedance:
Can be used with single-ended 50 loads or differential 100 loads.
Delivered bias voltages are halved with single-ended loads.
IQ bias voltages: Independent I and Q channel bias voltages settable within the range ±3 V.
Bias voltage: Resolution: 1.0 mV nominal
Accuracy: ±2% ±4 mV max, ±1% ±2 mV typical
Offset: see IQ bias voltages above
Differential offset voltage: Range: ±300 mV
Resolution: 100 µV nominal
Accuracy: ±2% ±3.3 mV max, ±1% ±0.7 mV typical
Level mode: Variable IQ signal level over 45 dB range
Differential signal balance: Typically 0.15 dB at 10 MHz
IQ channel balance: ±0.2 dB at 1 MHz
IQ level imbalance adjust: ±4 dB nominal, continuously variable
IQ signal amplitude: 22.4 mV to 4 V pk-pk per channel
IQ signal amplitude accuracy: <2% at 20 kHz,typically 1.5%, excludes termination errors
Baseband purity (2 V p-p set
voltage at 1 MHz):
2
nd
harmonic: 70 dBc
3
rd
harmonic: 65 dBc
IMD: 70 dBc (100 kHz tone spacing at 1 MHz)
REMOTE CONTROL
Ethernet: All signal generator parameters except the supply switch are remotely programmable.
The following LAN protocols are supported:
VXI-11
Telephone Network (TELNET)
File Transfer Protocol (FTP) (software upgrades only).
RS-232: All functions except the supply switch are remotely programmable.
Can be used for upgrading the firmware without removing the instrument’s covers.
GPIB: All functions except the supply switch are remotely programmable.
Capabilities: Designed in accordance with IEEE 488.2.
Interface functions: SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E2
ELECTROMAGNETIC
COMPATIBILITY
Conforms to the protection requirements of Council Directive 89/336/EEC.
Conforms with the limits specified in the following standards:
IEC/EN 61326-1 : 1997 + A1 : 1998 + A2 :2001
Emission: Class B.
Immunity: Table 1 and Performance Criterion B.
SAFETY
Conforms with the requirements of EEC Council Directive 73/23/EEC (as amended) and
the product safety standard IEC/EN 61010-1 : 2001 + C1 : 2002 + C2 : 2003 for Class 1
portable equipment, for use in a Pollution Degree 2 environment. The instrument is
designed to operate from an Installation Category 2 supply.
RATED RANGE OF USE
MIL-T-28800E Class 5
Temperature: 0 to 50°C (32 to 122°F).
Humidity: 45%, 0 to 50°C (32 to 122°F)
95%, 30 to 40°C (86 to 104°F)
Altitude: 700 mbar, 3050 m (10 000 ft)