EasyManua.ls Logo

Agilent Technologies 1000B Series - High Frequency Compensation; Figure 5 High Frequency Probe Compensation

Agilent Technologies 1000B Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started 1
Agilent 1000B Series Oscilloscopes User’s Guide 27
High Frequency Compensation
For the supplied passive probes:
1 Using the BNC adapter, connect the probe to a square wave generator.
2 Set the square wave generator to a frequency of 1 MHz, an amplitude of
3 Vp- p, and an output termination of 50Ω.
3 Press the [Auto Scale] front panel key.
4 If waveform does not appear like the Correctly Compensated waveform
shown in Figure 5, then use a nonmetallic tool to adjust the 2 high
frequency compensation adjustments on the probe for the flattest
square wave possible.
Figure 5 High Frequency Probe Compensation
High frequency compensation adjustments
Over Compensated
Under Compensated
Correctly Compensated

Table of Contents

Related product manuals