3Capturing Data
96 Agilent 1000B Series Oscilloscopes User’s Guide
To specify a trigger holdoff
Trigger holdoff can be used to stabilize a waveform. The holdoff time is
the oscilloscope's waiting period before starting a new trigger. The
oscilloscope will not trigger until the holdoff time has expired.
To specify a trigger holdoff:
1 Press [Menu].
2 In the Trigger menu, press Set Up.
3 In the Set Up menu, press Holdoff and turn the entry knob to adjust
the holdoff setting.
To reset the trigger holdoff
1 In the Set Up menu, select the Holdoff Reset menu item to return the
trigger holdoff setting to the 100 ns minimum value.
Figure 38 Trigger Holdoff