Agilent 1200 Series Autosampler Reference Manual 115
Repairing the Autosampler 4
3 Insert the new needle-seat assembly. Press the seat firmly
into position.
4 Connect the seat-capillary fitting to port 5 of the injection
valve.
5 Use “Down” to position the needle approximately 2 mm
above the seat
6 Ensure the needle is aligned with the seat. If required,
bend the needle slightly until the needle is aligned
correctly.