EasyManua.ls Logo

Agilent Technologies 1200 series - Page 87

Agilent Technologies 1200 series
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1200 Series Autosampler Reference Manual 87
Troubleshooting and Test Functions 3
Using the Change Needle Function
1 Ensure the front cover is installed.
2 Select “Start” (“Change”) to move the needle arm to the maintenance
position.
3 Remove the front cover.
4 Exchange the needle or needle seat (see “Needle-Seat Assembly” on
page 114 and “Needle Assembly” on page 111).
5 Replace the front cover.
6 Select “End” (“Done”) to complete the procedure.
NOTE
Do not remove the front cover until the needle arm is in its maintenance position. Removing
the cover while the needle arm is activated may lock up the system.

Table of Contents

Other manuals for Agilent Technologies 1200 series

Related product manuals