EasyManua.ls Logo

Agilent Technologies 6890 Series - Page 127

Agilent Technologies 6890 Series
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
115
Part 1. Capillary columns Columns and Traps
Procedure: Installing capillary columns in the ECD and µ-ECD
2. Install a 1/4-inch nut and graphitized-Vespel ferrule on the
adapter.
3. Prepare the column. See page 101 for instructions.
4. If you are using an indented liner (required for µ-ECD):
If the column id is 200 µm or more, push the column into the adapter until
it stops at the indentation. Pull it back 1 to 2 mm and tighten the column
nut firmly.
If the id is less than 200 µm, mark the column 70±1 mm from the end. Insert
column and nut into the adapter with the mark at the rear of the column
nut, and tighten the column nut firmly.
1/4-inch graphitized Vespel ferrule
1/4-inch nut
Mark
70 +
1 mm

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Related product manuals