71
Part 1. Electronic Pneumatic Control (EPC) Flow and Pressure Control
Maintaining EPC calibration
Procedure: Zeroing flow and pressure sensors
or
3. Scroll to a zero line and press [Info]
To zero a flow or pressure sensor in a specific module
1. Press [Options], scroll to Calibration, and press [Enter]
2. Scroll to the module to be zeroed and press [Enter]
Note: After zeroing or flow calibration,
the
Factory Calibration
line is replaced by the time and date of the
recalibration.
To restore the Factory Calibration, select
the time and date line and press [Delete].
This destroys the user calibration.
5. To cancel, press [Clear]
6. To zero, verify that the supply gas lines are disconnected from all modules.
7. Press [On] to zero or [Clear] to cancel.
4. To cancel, press [Clear]
5. To zero pressure, verify that the
supply gas line is not connected.
6. Press [On] to zero or [Clear] to cancel.
4. To cancel, press [Clear]
5. To zero flow, verify that the carrier gas
is connected and is turned on.
6. Press [On] to zero or [Clear] to cancel.
To zero all pressure sensors in all modules
1. Press [Options], scroll to Diagnostics, and press [Enter]
2. Scroll to
Electronics and press [Enter]
3. Scroll to
Pneumatics Board and press [Enter]
4. Scroll to
Zero all p sensors and press [Info]

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals