Messages Part 3. Warning messages
252
Part 3. Warning messages
Table 27 lists the Warning messages for the GC. Most require Agilent service
i n te r ve n ti o n . T ho s e th a t u s e rs c an c or r ec t a re i n d ic a te d al o n g w i th t he c or r e ct i ve
procedures.
Table 27 Warning Messages
Warning no. Status message Popup message Run log entry Comments
100 Oven sensor missing Oven sensor missing
101 Invalid heater power Invalid heater power for
front detector, inlet, and
aux 1
If using an MSD, make
sure the Aux zone is
configured for an MSD
transfer line..
102 Invalid heater power Invalid heater power for
front detector, inlet, and
aux 2
103 Sig 1 buffer full Sig 1 buffer full Possible data loss:
Sig 1 buffer full
See page 255.
104 Sig 2 buffer full Sig 2 buffer full Possible data loss:
Sig 2 buffer full
See page 255.
105 Analog out data loss Analog out data loss Possible data loss:
Analog out data loss
Contact Agilent service.
106 Signal data loss Non-recoverable data loss.
Data corrupt.
Possible data loss:
Signal data loss
Contact Agilent service.
107 F det config changed Front det: config changed,
method defaulted
Correct the method to
match your hardware.
108 B det config changed Back det: config changed,
method defaulted
Correct the method to
match your hardware.
109 F inl config changed Front inlet: config changed,
method defaulted
Correct the method to
match your hardware.
110 B inl config changed Back inlet: config changed,
method defaulted
Correct method to
match your hardware.
111 Col 1 config changed Column 1: config changed,
method defaulted
Correct method to
match your hardware.