Columns and Traps Part 5. Conditioning chemical traps
140
Part 5. Conditioning chemical traps
If your traps are preconditioned, you will not need to perform a conditioning
procedure before using them. However, all traps need regeneration periodically,
for example after using one to four cylinders of gas, or if gases of the highest
purity were not used. You can recondition Agilent moisture and activated
charcoal traps. Agilent oxygen traps cannot be reconditioned; you must replace
them if they become contaminated. Follow the manufacturer’s instructions for
reconditioning traps.
The molecular sieve and activated charcoal traps can also be repacked.
Instructions for repacking traps are shipped with each trap.
Table 18 Ordering Information for Agilent Traps
Item Part no.
Moisture trap (packed with Molecular Sieve 5A, 45/60 mesh) 5060-9077
Conditioned moisture trap (packed with preconditioned
Molecular Sieve 5A, 45/60 mesh)
5060-9084
Activated charcoal trap 5060-9096
Molecular Sieve 5A (100 grams, 45/60 mesh) 5080-6759
Activated charcoal (100 grams, 30/60 mesh) 5080-6751
Cap for ends of traps, 1/8-inch, 6 per package 5180-4124
Reducer trap fittings 5062-3502

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals