Flow and Pressure Control Part 1. Electronic Pneumatic Control (EPC)
Enter the rest of the inlet parameters
60
Procedure: Setting the rest of the inlet parameters
1. Press [Front Inlet] or [Back Inlet].
2. Scroll to the Mode line.
3. Press [Mode/Type] to see the inlet mode menu.
4. Move the cursor to the inlet mode you want. Press [Enter]. The inlet table
may change, depending on your choice. The possibilities for the defined-
column case are shown on the next page.
5. Scroll to Temp. Type the temperature you want. Press [Enter].
6. If you selected Split, and if the column is defined, you may enter the split
ratio directly.
For details on the inlet parameters, see the Inlets volume.
Press [Mode/Type] to see
the
Inlet Mode
menu.