EasyManua.ls Logo

Agilent Technologies 6890 Series - Ferrules and O-Rings for Glass Packed Columns

Agilent Technologies 6890 Series
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
135
Part 3. Packed glass columns Columns and Traps
Ferrules and O-rings for glass packed columns
Ferrules and O-rings for glass packed columns
Table 17 lists ferrules and O-rings used with glass packed columns. Consult the
Agilent Catalog for consumables and supplies for a more complete listing.
Ferrules that are prepared improperly cause leaks and contamination. To avoid
problems, place graphitized Vespel ferrules in a petri dish and bake in the GC
oven at 250 to 300°C for 30 minutes before use to remove organic compounds
absorbed by the graphite. Leave a petri dish of assorted ferrules in the GC oven
to ensure a clean supply.
Table 17 Glass Packed Columns Consumables
Item* Typical use Part no.
1/4-inch graphitized Vespel ferrule,
pkg of 10
Inlet/detector liners,
1/4-inch glass packed
columns
5080-8774
Silicone O-ring, 6.0-mm 1/4-inch glass packed
columns
0905-0322
* O-ring and ferrules ids

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Related product manuals