The Keyboard and Display
Default parameters
40
Default parameters
The GC software supplies default values for most parameters if you do not
specify them. These values are reasonable operating parameters for inlets and
detectors. Once you change a parameter, the default value for that parameter is
erased.
At some time, you may find it desirable to reload the default parameters. Doing
this erases all current parameters except any methods you have stored and
replaces them with the default set.
Procedure: Loading the default parameters
1. Press [Method]
2. Scroll to the Set default method line and press [Enter].
3. This message will appear:
4. Press [Enter] to load the default parameters.