EasyManua.ls Logo

Agilent Technologies 6890 Series - Column Calibration Procedures; Procedure: Estimate the Actual Column Length or Diameter from an Elution Time

Agilent Technologies 6890 Series
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Columns and Traps Part 6. Calibrating your capillary column (optional)
Column calibration procedures
142
Column calibration procedures
These procedures are described below using Column 1 as an example.
Procedure: Estimate the actual column length or diameter from an elution time
1. Set oven ramp 1 to 0.00, then verify that the column is defined. For more
information, see "Setting up an isothermal run" on page 87 or "Configure the
column" on page 48.
2. Perform a run using an unretained compound and record the elution time.
3. Press [Options]. Scroll to Calibration and press [Enter].
4. From the calibration list, select Column 1 or Column 2 and press [Enter].
5. The GC displays the current calibration mode for the column. In this exam-
ple, the column is uncalibrated.
Select the column to calibrate

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Related product manuals