The Keyboard and Display
Miscellaneous keys
32
3. Enter Oven temp and Column pres.
The Post Run LED on the status board is lit during a Post Run.
If you press [Time] while in a Post Run, you can view the amount of time
remaining.
[Run Log]
Deviations from the planned method (including keyboard intervention) during
the most recent run are listed in the run log table. Up to 50 run log entries can
be stored. The run log information can be used for Good Laboratory Practice
(GLP) standards. The run log can be uploaded to a workstation or printed out
on an integrator.
The Run Log LED is lit if there are any entries in the run log for the run in progress.
The run log is cleared at the start of a new run.
The instrument is now programmed to
maintain the oven temperature at 250°C
for 10 minutes following a run, with
column 2 head pressure at 15.0 psi.
Press [Run Log]

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals