253
Part 3. Warning messages Messages
Table 27, continued
Warning no. Status message Popup message Run log entry Comments
112 Col 2 config changed Column 2: config changed,
method defaulted
Correct method to
match your hardware.
113 Aux 3 method changed Aux 3 config changed
Method defaulted
Correct method to
match your hardware.
114 Aux 4 method changed Aux 4 config changed
Method defaulted
Correct method to
match your hardware.
115 Aux 5 method changed Aux 5 config changed
Method defaulted
Correct method to
match your hardware.
116 Log overflow Run log capacity is 50
entries.
117 F inl calib deleted F inl calib deleted
Inlet module is returned
to default calibration.
118 B inl calib deleted B inl calib deleted
119 F det calib deleted F det calib deleted Detector module is
returned to default
calibration.
120 B det calib deleted B det calib deleted
121 P aux calib deleted P aux calib deleted Module is returned to
default calibration.
122 Comm data overrun Host communications:
data overrun
Possible data loss:
Comm data overrun
Contact Agilent service.
123 Comm data error Host communications:
data error
Possible data loss:
Comm data error
Contact Agilent service.
124 Comm abnormal break Host communications:
abnormal break
Possible data loss:
Comm abnormal break
Check connection.
125 Sampler data overrun Sampler communications:
data overrun
Possible data loss:
Sampler data overrun
Check your sampler
settings. Contact
Agilent service.
126 Sampler data error Sampler communications:
data error
Possible data loss:
Sampler data error
Check your sampler
settings. Contact
Agilent service.
127 Sampler abnormal com Sampler communications:
abnormal break
Possible data loss:
Sampler abnormal com
Check connection.