Messages Part 3. Warning messages
254
Table 27, continued
Warning no. Status message Popup message Run log entry Comments
128 F inl flow cal fail Front inlet flow sensor
auto zero calib failed.
Contact Agilent service.
129 B inl flow cal fail Back inlet flow sensor
auto zero calib failed.
Contact Agilent service.
130 Aux 1 cryo disabled Aux 1 & front inlet on
same cryo valve drive:
aux1 disabled
Reconfigure aux or inlet
cryo drive.
131 Aux 2 cryo disabled Aux 2 & back inlet on
same cryo valve drive:
aux2 disabled
Reconfigure Aux or inlet
cryo drive.
132 Chgd Col 1 Init time to
###.## ; avoids Sampling
End problem
For Volatiles interface, a
setpoint conflicted with
the Sampling End time
parameter. Check your
method. See the Inlets
volume for more
information.
133 Chgd Col 2 Init time to
###.## ; avoids Sampling
End problem
134 Chgd FI Saver time to
###.## ; avoids Sampling
End problem
135 Chgd BI Saver time to
###.## ; avoids Sampling
End problem
136 Chgd FI Purge time to
###.## ; avoids Sampling
End problem
137 Chgd BI Purge time to
###.## ; avoids Sampling
End problem