Columns and Traps Part 3. Packed glass columns
Overview: Installing glass packed columns
132
Table 16 Installing Glass Packed Columns
Inlet or
detector
Where to
install
Column
configuration Comments
Purged-packed
inlet
Inlet fitting (no
liner installed)
or 1/4-inch liner*
A or B, depending on
the detector
C (works with all
detectors)
Allow at least 50 mm of empty column
to prevent an inserted syringe needle
from contacting either the glass wool
plug or column packing.
Adaptable
NPD, FID, or FPD
Cannot use with
capillary optimized
detector
Detector fitting A Remove adapter, if installed.
There must be at least 40 mm of empty
column to prevent the bottom end of the
jet from touching either column packing
or the glass wool plug.
ECD Detector fitting A Remove capillary column adapter, if
installed.
TCD 1/4-inch adapter
(Part no.
G1532-20710)
B Instructions for installing adapters are
on page 128.
* See the Inlets volume for instructions on installing liners

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals