SCPI COMMAND SUMMARY
be used, there is really only a single value for all the markers.
*RST value is 2 dB.
MARKer[n]:AOFF
Sets all the markers to
OFF
at once. While [n] may be used, there is
really only a single switch for all the markers.
MARKer[n]:DELTa? <num>,<num>
This query returns the dierence in frequency between the two
specied marker numb ers.
MARKer[n]:FREQuency <num>[freq suffix]|MAXimum|MINimum
MARKer[n]:FREQuency? [MAXimum|MINimum]
Sets and queries the sp ecied marker frequency (marker number
one is the default if [n] is not sp ecied). The value is interpreted
dierently based on the value of the marker mo de.
MARKer[n]:MODE How the frequency of the marker is determined.
FREQuency Absolute frequency is used. The limits are
conned to the present START and STOP
frequency limits.
DELTa The value is sp ecied with resp ect to the
reference marker. (MARKer[n]:REFerence)
The *RST values are the same as the
FREQ:CENT
command *RST
value.
MARKer[n]:MODE FREQuency|DELTa
MARKer[n]:MODE?
Sets and queries the mo de of the specied mark
er. Setting one
marker to delta turns all other marker mo des to frequency. If [n] is
not sp ecied, the default is one. *RST value is
FREQuency
.
MARKer[n]:REFerence <n>
MARKer[n]:REFerence?
Sets and queries which marker is the reference marker for use in
the delta mo de. While [n] may b e used, there is really only a
single reference for all the markers.
MARKer1:REFerence 5;
and
MARKer2:REFerence 5;
both set marker 5 as the reference.
MARKer[n][:STATe] ON|OFF|1|0
MARKer[n][:STATe]?
The state of the sp ecied marker is set and queried (marker number
one if [n] is not sp ecied). The *RST value for all markers is
OFF
.
MEASure:AM?
A query-only command that causes the mo dulating AM signal to be
S-40 Operating and Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620B and is the answer not in the manual?

Agilent Technologies 83620B Specifications

General IconGeneral
BrandAgilent Technologies
Model83620B
CategoryInverter
LanguageEnglish

Related product manuals