EasyManuals Logo
Home>Agilent Technologies>Test Equipment>InfiniiVision 2000 X-Series

Agilent Technologies InfiniiVision 2000 X-Series User Manual

Agilent Technologies InfiniiVision 2000 X-Series
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #202 background imageLoading...
Page #202 background image
202 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
15 Digital Voltmeter
The DVM makes accurate RMS measurements when the signal frequency is
between 20 Hz and 100 kHz. When the signal frequency is outside this
range, "<BW Limit?" or ">BW Limit?" appears in the DVM display to
caution you about inaccurate RMS measurement results.
To use the digital voltmeter:
1 Press the [Analyze] key.
2 Press Features; then, select Digital Voltmeter.
3 Press Features again to enable the DVM measurements.
4 Press the Source softkey and turn the Entry knob to select the analog
channel on which digital voltmeter (DVM) measurements are made.
The selected channel does not have to be on (displaying a waveform) in
order for DVM measurements to be made.
5 Press the Mode softkey and turn the Entry knob to select the digital
voltmeter (DVM) mode:
AC RMS — displays the root- mean-square value of the acquired data,
with the DC component removed.
DC — displays the DC value of the acquired data.
DC RMS — displays the root- mean- square value of the acquired data.
Frequency — displays the frequency counter measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 2000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 2000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 2000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals