Rockwell Automation Publication 1441-UM001B-EN-P - September 2012 195
Multi-channel Measurements Chapter 6
7. Select Input Channel and press the Right arrow to display the input
channel choices.
8. Select the channel and press the Left arrow key.
See Connector A and Connector B Inputs on page 19
for complete
information on the connectors and related channels.
9. Change the other parameters as needed.
See Offroute User Parameters on page 156
for assistance.
You can also set the Acquisition parameter in the Instrument
Configuration screen to Single Shot or Continuous.
– In Single Shot mode, the unit halts when the data collector has settled
and the data is stable.
– In Continuous mode, the unit constantly analyzes and displays the
moving signals until you store the data by pressing the Enter key.
See Instrument Configuration Parameters on page 59
.
10. You are now ready to collect the dual channel measurement.
See Collect a Cross Channel Phase Measurement on page 188
.