Spectrum Analyzer Performance Verification 3-5 Single Side Band (SSB) Phase Noise
Model MS20xxB MM PN: 10580-00303 Rev. D 3-9
26. Press the Delta On/Off submenu button to turn On Delta.
27. Use the numeric keypad to enter 10 and then press the kHz submenu button.
28. Record the measured value in Appendix A.
• MS2034B – Table A-13, “Spectrum Analyzer SSB Phase Noise” on page A-6
• MS2035B – Table A-38, “Spectrum Analyzer SSB Phase Noise” on page A-18
29. Subtract 30 dB from the measured value and verify that the result is better than –100 dBc/Hz (for
10 kHz offset) or –105 dBc/Hz (for 100 kHz offset) or –115 dBc/Hz (for 1 MHz offset) and record the
calculated values in Table A-13 for MS2034B or Table A-38 for MS2035B in Appendix A.
For example: –70 dBc (measured) –30 dB = –100 dBc/Hz
30. Repeat Step 15 through Step 29 for 100 kHz offset (set Span to 220 kHz) and 1 MHz offset (set Span to
2.04 MHz). Record the measured values and calculated values in Table A-13 for MS2034B or Table A-38
for MS2035B in Appendix A.