3-10 Third Order Intercept (TOI) Spectrum Analyzer Performance Verification
3-32 PN: 10580-00303 Rev. D Model MS20xxB MM
8. On the MG3692x Synthesized Signal Generator #2, turn the RF On, and adjust the RF amplitude so that
–20 dBm appears at the splitter output (approximately +16 dBm).
9. Turn Off the RF of the MG3692x Synthesized Signal Generator #2.
10. Set the MS203XB to Spectrum Analyzer mode and then Preset the instrument.
11. Press the Shift key, the Sweep (3) key, Sweep Mode submenu key and then the Performance submenu
key.
12. Press the Back button and press the Detection submenu button and then the RMS/AVG submenu button.
13. Set the MS203xB as follows:
Center Frequency: 800.151 MHz
Span: 100 Hz
RBW: 30 Hz
VBW: 1Hz
Reference Level: –15 dBm
Atten Lvl: 0dB
Preamp: Off