A7 Plus OPERATOR'S MANUAL
EN
15
Cleaning the dentist's board.
Clean the dentist's board using a suitable product (see paragraph 1.4).
NOTE for CONTINENTAL version instrument boards:
instrument holder ( x ) can be removed to make cleaning
operations easier; to remove it, simply extract it from its seat as
it is fixed with magnets.
The silicone instrument holder ( u ) can also be sterilised in an
autoclave up to 135°C.
Removable instrument cords.
All instruments are provided with removable cords to make cleaning
operations easier.
NOTE for CONTINENTAL version instrument boards:
to remove cords, first remove instrument holder, then loosen
the relevant plastic ring nuts.
NOTE for INTERNATIONAL version boards:
to remove cords, simply loosen the relevant plastic ring nuts
positioned under board.
WARNING:
Before removing instrument cords, turn operating unit
off.
After having turned operating unit off, drain syringe ducts
by pressing the relevant air and water buttons directly on
bowl until all spray water has flowed out.
TURBINE, MICROMOTOR and SCALER instrument cords
contain water; it is therefore recommended to remove
cord by keeping the handpiece side end on the bowl.
When refitting a cord, make sure that electric contacts are
dry and that the plastic ring nut is well tightened.
Each cord must be refitted only and exclusively inside the
corresponding instrument holder.
Clean the instrument cord using a suitable product (see Paragraph 1.4).
WARNING:
Instrument cords are NOT designed for sterilization in
autoclave or by being cold soaked in solution.