Chapter 19-6
6. Setting to Defaults
If you want to resume the default setup of the factory, select System
SetupSet To Defaultsthe confirm window appears, select OK.
19.3.6 Wifi Setup
You can connect the EMS to the central monitoring system according to
the following operation:
Press the button to enter the main menu. Select
System Setup
WIFI State to enter the Wifi state window.You can open/close the Wifi
function and check the related configuration message.
Open/CloseWifi
: Open or close Wifi function. After closing the
monitor, the Wifi state before closing will be saved. If the Wifi function is
open before closing, the monitor will connect to the central monitoring system
automatically when it’s open again.
WifiState: Show the current state of the wireless connection. There are
several state as follows:
Scanning: Scanning and trying to connect to AP;
Connecting CMS: Connected to AP and trying to connect to the central
monitoring system.
Connected CMS: Connected to the central monitoring system.
Error: Wifi module error. Contact after-sales personel please.
After connecting to the central monitoring system successfully, the
icon will appear in the top right corner of the screen. The icon will
change according to the strength of the signal.
SSID /LocalIP : Must be in accord with the settings of SSID and
segment of the monitor. The IP address cant be in accord with other machines
in the segment.
LocalMachineNo: Bed number displayed in the central monitor, and it
can’t be in accord with other machines.
FrequencyChannel : The preferred search frequency band. You dont
need to modify generally.
Caution:
Only your EMS module matches the Wifi function , the monitor has
the Wifi function.