Patient monitor User’s manual
Chapter 19-21
1. NIBP Displaying
2. Setting NIBP
Enter the main menu, Select
Module Setup
NIBP Setup
, a SpO
2
setup window appears. In the window you can set SpO
2
measurement.
a. Setting Unit
Select
Unit
, upturn or downturn to choose the unit you need. You can
choose betweenmmHgandkPa.
b. Selecting Time Interval
When
Mode
is set to
Auto
, select
Measure Mode
to choose an
appropriate periodic measurement time.
c. Setting Initial Press
SelectInflation, upturn or downturn to choose an appropriate initial cuff
inflation pressure according to patient category and requirement.
d. Setting Hypertension Mode
Select Hypertension Mode , upturn or downturn to choose an
appropriate setup. You can choose
ON
or
OFF
.
e. Setting Alarm Source
Select ALM Source to choose an appropriate setup. You can choose
from
SYS&DIA&MEAN
,
DIA
,
SYS
,
MEAN
,SYS&MEAN
DIA&MEANandSYS&DIA.
f. Resetting
Select
Reset
to restore the inflation value of blood pressure pump to the
initial value. When BP pump works abnormally but no prompt is
displayed, you may press Reset to check BP pump and to automatically
restore the abnormal working of BP pump caused accidentally.
3. Starting and Stopping NIBP Measurement
Sys
Unit
Dia
Mean