Patient monitor User’s manual
Chapter 26-6
non-sterile devices. Do not autoclave the devices as this will damage
Never sterilize or immerse the IRMA sensor in liquid.
7. Sidestream AG module
The ISA sidestream module should be cleaned on a regular basis.
Use a cloth moistened with max 70% ethanol or isopropyl alcohol to clean
To prevent cleaning liquids and dust from entering the ISA sidestream
module through its inlet port, keep the Nomoline sampling line connected while
The Nomoline sampling lines are non-sterile devices. To avoid damage,
do not autoclave any part of the sampling line.
Never sterilize or immerse the ISA sidestream module in liquid.
Replacement of sampling line
The Nomoline sampling line is non-reusable.
Every two weeks or whenever "Sampling line clogged" appears, whichever
comes first, the Nomoline sampling line should be replaced.
8. ICG cable, CSM cable, C. O. Interface Cable, BIS cable and Injection
Temperature Probe.
The method for the maintenance of ICG cable, please refer to that of ECG
Warning: Never use EtO or formaldehyde for disinfection.
Caution: Sterilization maybe do some harm to monitor, it is suggested
that the sterilization be performed only when your hospital
considers it necessary in line with maintenance plan. Clean
equipment before sterilization.
Note: The electric schematic and element list can only be offered to
the eligible service center or personnel.