Patient monitor User’s manual
27
HAL
4%
ENF, ISO, SEV
5%
DES
15%
Xe (Xenon)
80%
-10% of reading
1)
He (Helium)
50%
-6% of reading
1)
Metered dose inhaler
Propellants
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol)
0.3%
C
3
H
7
OH
(Isopropanol)
0.5%
CH
3
COCH
3
(Acetone)
1%
CH
4
(Methane)
3%
CO (Carbon
monoxide)
1%
NO (Nitrogen
monoxide)
0.02%
O
2
100%
Note 1: Interference at indicated gas level. For example, 50 % Helium typically decreases
the CO2 readings by 6%. This means that if measuring on a mixture containing 5.0 % CO2
and 50 % Helium, the actual measured CO2 concentration will typically be (1-0.06) * 5.0
% = 4.7 % CO2.
A.6.9 ICG
Measurement way
Measurement of thoracic electrical bioimpedance
Measurement range
HR: 40 bpm to 250 bpm
SV: 5 mL to 250 mL
SI: 5 mL/m
2
to 125 mL/m
2
C.O. : 1.4 L/min to 15 L/min
TFC: 15 /kΩ to 143 /kΩ
Accuracy
HR: ±2 bpm
SV: unspecified
C.O. : unspecified
Alarm range
C.I.: 0.0 L/min/m
2
to 15.0 L/min/m
2
, high/low limit can
be adjusted continuously.
TFC: 10 /kΩ to 150 /kΩ, high/low limit can be adjusted
continuously.