Patient monitor User’s manual
28
ICG (option)
Measurement way
Measurement of thoracic electrical bioimpedance
Measurement range
HR: 40 bpm to 250 bpm
SV: 0 mL to 250 mL
SI: 0 mL/m
2
to 125 mL/m
2
C.O. : 0 L/min to 30 L/min
TFC: 5 /kΩ to 150 /kΩ
Accuracy
HR: ±10% bpm
SV: ±15%
C.O. : ±15%
Alarm range
C.I.: 1.5 L/min/m
2
to 5.0 L/min/m
2
, high/low limit can be
adjusted continuously.
TFC: 15 /kΩ to 60 /kΩ, high/low limit can be adjusted
continuously.
Recovery time of
module after
defibrillation
The ICG module is protected against cardiac
defibrillations and will return to the standard operation
within 10s after discharge.
A.6.10 C.O.
Measurement Mode
Thermal dilution method
Measurement Wave
Thermal dilution curve
Measurement
parameters
C.O., TB, TI, C.I.
Measurement Range
C.O. : 0.1 L/min ~ 20 L/min
TB: 23.0 ~ 43.0
TI: -1.0 ~ 27.0
Resolution
C.O.: 0.1 L/min
TB: 0.1
TI: 0.1
Precision
C.O.: 2% SD
TB.TI: ±0.1
TB Alarm range
23.0~43.0
, high/low limit can be adjusted continuously
A.6.11 CSM
EEG sensitivity
±400µV