Patient monitor Users manual
59
Search pulse
SpO
2
module is searching for pulse
No alarm
Temp
Physiological alarm:
Alarm message
Cause
Level
<T1 label > high
T1measuring value is above high
alarm limit
Medium
User-selectable
< T1 label > low
T1 measuring value is below low
alarm limit
<T2 label > high
T2 measuring value is above high
alarm limit
<T2 label > low
T2 measuring value is below low
alarm limit
Technical alarm:
Alarm messages
Cause
Level
<T1 label > off
T1 sensor may be disconnected from
user or monitor
Low
<T2 label > off
T2 sensor may be disconnected from
user or monitor
<T1 label > Over Range
T1 is over measuring high range
<T1 label > Below Range
T1 is below measuring low range
<T2 label > Over Range
T2 is over measuring high range
<T2 label > Below Range
T2 is below measuring low range
Temp error
Temp calibration failure
NIBP
Physiological alarm:
Alarm messages
Cause
Level
NIBP SYS High
NIBP SYS measuring value is
above high alarm limit
Medium
User-selectable
NIBP SYS Low
NIBP SYS measuring value is
below low alarm limit
NIBP DIA High
NIBP DIA measuring value is
above high alarm limit
NIBP DIA Low
NIBP DIA measuring value is
below low alarm limit
NIBP MAP High
NIBP MAP measuring value is
above high alarm limit