Patient monitor User’s manual
60
NIBP MAP measuring value is
below low alarm limit
Technical alarm(displayed in NIBP parameter displaying area):
Transducer or other hardware failure.
a. Cuff is completely unwrapped.
b. The cuff is not connected.
c. Adult cuff used in neonate mode.
Air leak in pneumatics, hose, or cuff.
Unable to maintain stable cuff pressure,
e.g. kinked hose.
a. Very weak patient signal due to a
loosely wrapped cuff.
b. The pulse of patient is too weak.
Measurement range exceeds module
specification.
a. Too many retries due to interference of
motion artifact.
b. Signal is too noisy during
measurement, e.g. patient has severe
tremor.
c. Irregular pulse rate, e.g. arrhythmia.
Cuff pressure exceeds the specified high
safety limit. Adult: 297 mmHg, pediatric:
252 mmHg
,
Neonate: 147 mmHg
。
Maybe
caused by quick extrusion or striking
cuff.
Large motion artifact that saturates the
BP amplifier’s amplitude handing
capability.
Module reports Air Leakage failure while
in the Pneumatic Test mode.
Module occurs abnormal processor event.
Measurement took more than 120
seconds in adult, 90 seconds in neonate
mode.
Neonate cuff used in adult mode.
Prompt messages(displayed in NIBP parameter displaying area):
NIBP measurement module is resetting