Patient monitor Users manual
61
OverPress Testing
NIBP is testing Over-Pressure
Manometer Testing
NIBP is testing Manometer
Air Leakage Testing
NIBP is testing Air Leakage
IBP
Physiological alarm:
Alarm messages
Cause
Level
P1 SYS High
SYS measuring value of channel 1 is
above high alarm limit
Medium
User-selectable
P1 SYS Low
SYS measuring value of channel 1 is
below low alarm limit
P1 DIA High
DIA measuring value of channel 1 is
above high alarm limit
P1 DIA Low
DIA measuring value of channel 1 is
below low alarm limit
P1 MEAN High
MEAN measuring value of channel 1 is
above high alarm limit
P1 MEAN Low
MEAN measuring value of channel 1 is
below low alarm limit
P2 SYS High
SYS measuring value of channel 2 is
above high alarm limit
P2 SYS Low
SYS measuring value of channel 2 is
below low alarm limit
P2 DIA High
DIA measuring value of channel 2 is
above high alarm limit
P2 SYS Low
DIA measuring value of channel 2 is
below low alarm limit
P2 MEAN High
MEAN measuring value of channel 2 is
above high alarm limit
Medium
User-selectable
P2 MEAN Low
MEAN measuring value of channel 2 is
below low alarm limit
Technical alarm:
Alarm messages
Cause
Level
P1 off
IBP sensor of channel 1 disconnect from
EMS
Low
P2 off
IBP sensor of channel 2 disconnect from
EMS