Appendix D Guidance and Manufacturer’s
Declaration of EMC
Guidance and manufacturers declaration electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration electromagnetic emission
The monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the monitor should assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The monitor uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class A
The monitor is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies building used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies