4.3.5 Sound on door closing
Line 14:1, Authorization level: C
When this function is activated, a signal is heard when the door is
being closed. The signal may be short (50 ms) or long (100 ms).
00 = no signal
01 = Signal 50 ms
02 = Signal 100 ms
Default value is 1.
4.3.6 Door type
Line 14:2, Authorization level: C
Changeover between manual and automatic door (option)
1 = Manual
2 = Automatic (option)
Default value is 1 for FD1800/FD1805 and 2 for FD1810.
4.3.7 IR activation time
Line 14:3, Authorization level: C
Setting of time, in seconds, that the IR sensor must be active for
the door to open (option).
±
Default value is 1.0. The value does not affect closing.
4.3.8 Acoustic alarm
Line 15, Authorization level: C
When this function is activated, a signal sounds at the end of the
program and when a error that causes a error message on the
display occurs.
00 = no signal
01 = signal
Default value is 00.
4.3.9 Feedback
Line 16, Authorization level: Factory (testing)
The feedback function is activated and adjusted.
01 = Activation of feedback
02 = Interval for feedback
4.3.10 Temperature scale
Line 17, Authorization level: C
Conversion between °C and °F. The display shows temperatures in
the selected unit.
1 = °Celsius
2 = °Fahrenheit
Default value is 1.