EasyManua.ls Logo

Arjo Tornado - Error and Program Statistics; Error Statistics

Arjo Tornado
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
4.10.5 Addition of softened water
Line 54, Authorization level: C
Specify whether soft water is to be added and if so where.
7KHWLPHIRUZKLFKWKHZDWHUZLOOEHDGGHGLVVSHFL¿HGDWOLQH
00 = No soft water is to be added.
01 = Soft water is to be added in the tank.
02 = Soft water is to be added in the steam generator.
Default value is 00.
4.10.6 Time for soft water
Line 55, Authorization level: C
Length of time for which soft water is to be added if the answer to
line 54 was 01 or 02.
00 - 99 seconds in one-second steps.
4.10.7 Flushing time, cooling (internal)
Line 56, Authorization level: C
7KHFRROLQJSKDVHÀXVKHVIRUVHFRQGV,IPRUHFRROLQJLV
required, the time can be extended
E\±VHFRQGV
Default value is 0.0.
 3XPSWLPHÀXVKLQJWLPHLQWKHLQVWUXFWLRQ³3XOVHÀXVKLQJWROHYHO´
Line 57, Sub-line 1, For future use
 3DXVHWLPHLQWKHLQVWUXFWLRQ³3XOVHÀXVKLQJWROHYHO´
Line 57, Sub-line 2, For future use
 3XOVHVLQWKHLQVWUXFWLRQ³3XOVHÀXVKLQJWROHYHO´
Line 57, Sub-line 3, For future use
4.11 Error statistics, lines 60 - 63
4.11.1 Time of error
Line 60
Shows the number of seconds for which the program had been
running when the error occurred.
00 - 9999 seconds
This value cannot be changed.