4.5 Interval disinfection, line 31
4.5.1 Interval disinfection active
Line 31, Sub-line 01, Authorization level: C
Here you can activate or deactivate the interval disinfection
function.
00 = Not active
01 = Active
Default value is 00.
4.5.2 Interval disinfection programs
Line 31, Sub-line 02, Authorization level: C
Choose the program to be run on interval disinfection. The
SURJUDPVZKLFKDUHGHVFULEHGLQGHWDLOXQGHU³3URJUDPVHOHFWLRQ´
have the following numerical values:
00 = No program
01 = Economy program
02 = Normal program
03 = Intensive program
04 = Normal program + dosing
05 = Intensive program + dosing
Default value is 01.
4.5.3 Limit for interval disinfection
Line 31, Sub-line 03, Authorization level: C
Set the number of programs performed between interval
disinfections.
0 - 9999
Default value is 9999.
4.5.4 Processes remaining to interval disinfection
Line 31, sub-line 04
Number of programs to the next interval disinfection. The value
cannot be changed but it can be reset; see line 31, sub-line 05.
4.5.5 Reset interval disinfection
Line 31, Sub-line 05, Authorization level: BC
Zeroing of the counter of number of processes remaining to
interval disinfection; see line number 31, sub-line number 04.
01 = reset counter
4.6 Inspection request, line 32
This function is used to get a continuous maintenance service of
the machine. When the function is active, a message stating that
an inspection is due appears on the display after a certain number
of hours or program runs.