EasyManua.ls Logo

Arjo Tornado - Disinfection and Cooling Control

Arjo Tornado
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
4.8 Empty container alarm, line 34
/LQH6XEOLQHV±$XWKRUL]DWLRQOHYHO&
Here you can specify how the empty container alarms from the
various containers are managed.
The value consists of two or four digits:
00 = No alarm
01XX = 01 means that a warning will be given. The last two digits,
±VSHFLI\WKHQXPEHURISURFHVVHVWKDQFDQVWLOOEHUXQDIWHU
the warning.
02 = An error alarm will be given. The machine cannot run until the
alarm has been acted on.
The various sub-lines correspond to the following agents:
01 = Alarm for descaler.
01XX = Alarm for descaler, XX processes remaining. Default value
is 25.
02 = Alarm for chemical agent.
02XX = Alarm for chemical agent, XX processes remaining. Default
value is 0.
03 = Alarm for rinse-aid.
03XX = Alarm for rinse-aid, XX processes remaining. Default value
is 25.
04 = Alarm for process agent.
04XX = Alarm for process agent, XX process agent remaining.
Default value is 10.
05 = Alarm for spray agent.
05XX = Alarm for spray agent, XX processes remaining. Default
value is 10.
4.9 Disinfection, lines 35 - 49
4.9.1 Normal disinfection method
Line 35, For future use
4.9.2 Standby disinfection method
Line 36, For future use
4.9.3 Viewing mode
Line 37, Authorization level: C
Indicates what is shown in the display.
1 = Temperature
2 = A
0
value
3 = A
0
value / Temperature
4 = Remaining time
5 = Remaining time/Temperature
6 = Remaining time / A
0
value
7 = Remaining time / A
0
value / Temperature
Default value is 7.