44
4.9.4 Lower disinfection temperature
Line 38, Authorization level: C
The heating is switched on at this temperature.
80-95°C
Default value is 91.
4.9.5 Upper disinfection temperature
Line 39, Authorization level: C
The heating switches off at this temperature. The upper
disinfection temperature must be 2 degrees above the lower
disinfection temperature and above the disinfection temperature.
±&
Default value is 93.
4.9.6 Disinfection temperature
Line 40, Authorization level: C
When this temperature is reached, the disinfection time in line 41
starts to count down.
80-95°C
Default value is 91.
4.9.7 Disinfection time
Line 41. Authorization level: C
Time for thermal disinfection.
0 - 1,800 seconds in one-second steps.
Default value is 35.
4.9.8 Max preheating time
Line 42, Authorization level: C
Maximum time for preheating of the steam generator.
0 - 99 seconds in one-second steps.
Default value is 50.
4.9.9 Power sharing, steam generator
Line 43, Authorization level: C
The steam generator has three elements. When this function
is active, only two elements are switched on when the pump is
running. This only applies to single-phase machines.
00 = Not active
01 = Active
Default value is 01.