4.9.10 Cooling method
/LQH6XEOLQHV±$XWKRUL]DWLRQOHYHO&
44:01 = Choose a method for cooling the washed items after
disinfection.
0 = no cooling
1 = Internal cooling from tank (does not meet EN ISO 15883)
2 = Ventilator/external cooling with fan
3 = Extended cooling after process
Default value is 2.
44:02 = Time for cooling
±V
Default value is 120.
44:03 = Prolonged external cooling time (after process)
Prolonged time ventilator
±V
Default value is 120.
4.9.11 Handling self-disinfection
Line 45, Authorization level: C
This selection determines how self-disinfection is handled.
01 = Choice of how to handle self-disinfection.
02 = What is the earliest time that self-disinfection should begin
03 = What is the latest time that self-disinfection should begin
WARNING!
If the machine has not been used for more than
72 hours, it is necessary to run a program with an
empty chamber before using the machine to clean
goods.
4.9.12 Disinfection A0
Line 46, Authorization level: C
0 = Disinfection according to time and temperature on line 40 and
line 41.
60 = Disinfection to level A
0
60
600 = Disinfection to level A
0
600
Default value is 0.
4.9.13 Dosage, chemical agent
Line 48, Authorization level: C
The chemical agent is a substance that is dosed into the tank in
WKH¿QDOSKDVHRIWKHSURJUDP+HUH\RXFDQVSHFLI\WKHWLPHIRU
which the dosing pump will run.
0 - 99 seconds in one-second steps.
Default value is 70.