EasyManuals Logo

AVL F-FEM-CON User Manual

Default Icon
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Operation
38
F-FEM-CON — Users Guide
4.1.2
Acquisition Rate and Aliasing
The structure of the ADC ensures excellent data sampling. Sampling is done at a
frequency many times that of the parameterized sampling rate (oversampling)
and is reduced to the sampling rate by a digital filter.
Combined transfer function of the converter (analog and digital)
Frequency response for f
notch
=10 Hz (normalized digital filter response)
The frequency f
notch
is identical to the sampling rate. The filter cut-off frequency
is 0.262×f
notch
. The low pass filter is of type
Fig. 20
Fig. 21
sin x
x
3

Table of Contents