07-19-B1-729 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 8-35
Removal/Replacement Procedures Disassembly/Reassembly
d. The screws must be M4x10, with captive washers.
e. Loctite 425 threadlocker must be applied to each of the three
screws.
f. Torque the screws to 9 kgf-cm (7.8 in-lb).
2. If not already present, apply an upgrade label (Part no. 072601047)
between the feet on the front case half. Use indelible ink (permanent
marker) and mark box number 2.
3. Perform the Operational Checkout procedures in Chapter 10,
“Operational Checkout”.