07-19-B1-729 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 10-3
Operational Checkout Self Test
3. Spike the solution container and fully prime the set.
4. Remove all trapped air bubbles from all set components.
5. Hang the solution container such that the fluid level is a minimum of 27"
above midpoint of the pump.
Q Ensure that the tubing is clean and dry before placing it in the pump.
Q Ensure that the tubing is placed and seated properly in the guide
channel, pump mechanism, sensors and safety clamp.
Q Ensure that there is no slack in the tubing and that it is not kinked or
pinched before closing the pump door.
6. Place the distal end of the administration set in a container or sink to
dispose of pumped solution.
Self Test
1. With the device disconnected from AC power and powered OFF, press
and hold the
SILENCE key and then press the ON-OFF/CHARGE key for
either Pump 1 or Pump 2 to power it ON. Do not release the
SILENCE
key until instructed in step 3.
2. Check the software version on the Main Display. The word SOFT
displays, followed by the version number. Record the software version
number on the data sheet.
3. Release the SILENCE key and allow the self test to complete. Verify the
following occurs:
Note: If you miss any of the indications listed below during the self test, repeat the
self test as many times as required to observe the indications. To repeat the
self test turn the device off then back on.
Q BEGINNING OF SELF TEST:
• All LEDs turn on.
• All segments of the Message display for both pumps turn on,
along with all segments of the Main display for the pump that
was powered ON.
• All backlit icons are turned on except for the plug icon.
• The main buzzer will turn on and change tone while all
segments of the displays are on.
• The backup buzzer will turn on after the main buzzer turns off.
• When the backup buzzer turns off, all LEDs will turn off.