7 Automatic Test Modes
7-8 FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-B1-729
1. Turn the pump off.
2. Press and hold the CLEAR TOT VOL key and while pressing the numeric
key corresponding to the automatic test mode you wish to enter (1
through 0).
3. While performing step 2, press the ON-OFF/CHARGE key.
4. To exit any automatic test mode, press the ON-OFF/CHARGE key.
Automatic Test Mode 3: Manufacturing Test Mode
The initial display in this automatic mode is for a PRI RATE of 1999 mL/hr,
PRI VTBI of 3998 mL, SEC RATE of 500 mL/hr, and SEC VTBI of 100 mL.
Pressing the
TIME key changes the primary settings as follows:
In each combination, the primary and secondary rates and VTBI can be
manually altered.
The pump stops automatically one hour after the
PRI START key was
pressed. While the pump is running, the air sensor and the upstream and
downstream occlusion sensors are disabled.
This is an automatic test mode for testing the pump during manufacturing only.
This mode must not be used on patients because the downstream occlusion,
upstream occlusion and air sensors are disabled.
PRI RATE PRI VTBI SEC RATE SEC VTBI
999 1998 500 100
499 998 500 100
250 500 500 100
125 250 500 100
10 20 500 100